**YS-227Y Medical Protective Clothing Synthetic Blood Penetration Tester (4‑Station)**
**【Scope of Application】**
Used for high‑pressure penetration testing of test products by delivering high‑pressure output.
**【Applicable Standards】**
- GB 19083-2010 Technical requirements for medical protective masks
- YY 0469-2011 Medical surgical masks
- YY/T 0691-2008 Test method for resistance of medical face masks to penetration by
synthetic blood (fixed volume, horizontally projected)
- ISO 22509:2004; F1862/F1862M-2013
- ASTM F1671 Standard test method for resistance of protective clothing materials to
blood‑borne pathogen penetration using Phi‑X174 bacteriophage penetration as a test system
- ASTM F1670 Standard test method for resistance of protective clothing materials to
synthetic blood penetration
- ASTM F903 Standard test method for resistance of protective clothing materials to
liquid penetration
- ISO 16603:2004 Protective clothing against contact with blood and body fluids —
Determination of the resistance of protective clothing materials to penetration by
blood and body fluids — Test method using synthetic blood
- ISO 16604:2004 Protective clothing against contact with blood and body fluids —
Determination of resistance of protective clothing materials to penetration by
blood‑borne pathogens — Test method using Phi‑X174 bacteriophage
- YY/T 0699-2008 Test method for resistance of protective clothing materials to
penetration by liquids under pressure for liquid chemical protective equipment
- YY/T 0689-2008 Test method for resistance of protective clothing materials to
penetration by blood‑borne pathogens — Phi‑X174 bacteriophage test method
- YY/T 0700-2008 Test method for resistance of protective clothing materials to
penetration by blood and body fluids — Synthetic blood test method
- GB/T 19082-2009 Technical requirements for disposable medical protective clothing
**【Instrument Features】**
1. Equipped with a silent air pump.
2. Equipped with a positive pressure sensor for real‑time pressure monitoring.
3. Fitted with a precision pressure regulating valve, allowing stepless pressure
adjustment and setting of technical parameters such as holding pressure and holding time.
4. Adjustable spray time with automated operation for convenience and speed.
5. Adopts a mask‑holding device for easy sampling and testing.
**【Technical Parameters】**
1. **Blood volume:** 2 mL / 60 mL test
2. **Control mode:** Manual control / push‑button control
3. **Adjustable pressure:** 16 kPa / 2 psi for high‑pressure output, accuracy 0.1 kPa
4. **Test pressure range:** 0–16 kPa (2 psi)
5. **Micro‑pressure gauge range:** 0–0.1 kPa
6. **Pressure resolution:** 0.1 kPa
7. **Time accuracy:** 0.1 s
8. **Test medium:** Synthetic blood (provided by the customer)
9. **Control method:** Manual / push‑button
10. **Blood volume:** 2 mL / 60 mL
**【Test Procedure】**
Secure the specimen in the specimen holder. At a distance of 30.5 cm from the center of the specimen,
spray 2 mL / 60 mL of synthetic blood (surface tension 0.042 ± 0.002 N/m) horizontally onto the target
area of the specimen through a needle with an inner diameter of 0.84 mm, at a pressure of
16 kPa / 2 psi. Remove the specimen and visually inspect it within 10 s / 60 s.