Scope of Application
【Operating Instructions】
[Scope of Application]Used for testing the photo-thermal properties of various fabrics and textile products. It uses a xenon lamp as the irradiation source, placing the specimen at a specified distance under a certain irradiance. The specimen's temperature rises due to the absorption of light energy, which is used to measure the photo-thermal properties of textiles.
[Compliance Standards]GB/T 18319-2019 "Textiles - Test method for photo-thermal properties"
[Instrument Features]
Large-screen color touch screen display and operation. Menu-driven operation with Chinese and English interfaces.
Equipped with an imported xenon lamp lighting system.
Equipped with high-precision imported temperature sensors.
Automatically measures and displays warm-up time, light time, dark time, xenon lamp irradiance, specimen temperature, and ambient temperature during the test.
Automatically records the temperature changes of the specimen and environment over time during the test. Automatically turns off the xenon lamp when the preset irradiation time is reached, and automatically calculates the maximum temperature rise and average temperature rise. The computer automatically draws the time-temperature curve.
Reports store test data, automatically calculate test maximum, minimum, average, mean square deviation, and coefficient of variation (CV%) values. Equipped with print and online interfaces.
[Technical Specifications]
Temperature Rise Test Range: 0~100℃; Resolution: 0.01℃
Average Temperature Rise Test Range: 0~100℃; Resolution: 0.01℃
Xenon Lamp: Spectral range (200~1100) nm; capable of producing (400±10) W/m² irradiance at a vertical distance of 400mm; irradiance is adjustable.
Temperature Sensor: Accuracy 0.1℃
Temperature Recorder: Capable of continuously recording temperature every 1min (recording interval setting range 5s~1min).
Irradiance Meter: Measurement range (0~2000) W/m²
Timer Range: Light time and cooling time setting range 0~9999min; Accuracy 1s
Vertical distance between specimen stage and xenon lamp: (400±5)mm. The temperature sensor is located at the center below the specimen and can be in full contact with it.
Dimensions: L 460mm, W 580mm, H 620mm
Weight: 42Kg
Power Supply: AC220V, 50Hz, 3.5KW (Requires matching 32A)