Scope of Application
【Operating Instructions】
**YG-WH Medical Device Packaging Physical Performance Tester**
**【Technical Parameters】**
1. **Three sensors as standard:** 0–5 kg, 0–25 kg, 0–50 kg
2. **Measurement units:** N, kg, lb
3. **Accuracy class:** Class 0.3 (0.5)
4. **Test force measurement range:** 2% – 100%
5. **Test force resolution:** 1/300,000 of the maximum test force
6. **Deformation measurement range:** 2% – 100%
7. **Relative error of deformation indication:** < ±1%
8. **Deformation resolution:** 1/500,000 of the maximum deformation
9. **Relative error of displacement indication:** < ±0.1%
10. **Data acquisition frequency:** 50 ms per reading
11. **Displacement resolution:** 0.001 mm
12. **Test stroke:** 1000 mm
13. **Test modes:**
- Opening force test
- Breaking force test
- Puncture force test
- Tensile/compression test
- Needle tube rigidity test
- Constant force test
- Constant displacement test
- Peel force test
- Piston sliding performance test
- Device body sealing test
14. **Standard configuration:** Main unit, plastic specimen test fixture, fabric specimen test
fixture, thread/filament specimen test fixture, syringe specimen test fixture, computer, testing software