Scope of Application
【Operating Instructions】
**YG-601Pro Leakage Current Tester**
**【Applicable Standards】**
Complies with the requirements for testing all leakage current items specified in GB 9706.1-2020 "Medical electrical equipment —
Part 1: General requirements for basic safety and essential
performance" for active medical device inspection, as well as all leakage current items specified in IEC 61010-1 (2001),
IEC 60950-1 (2005), and IEC 60335-1 (2001) + A1:2004 + A2:2006.
**【Technical Parameters】**
1. **Maximum rated current:** 20 A
2. Capable of performing leakage current testing without frequency weighting without connecting external equipment or devices; polarity switching without power interruption.
3. The device is equipped with external pin hardware interfaces for S1, S5, S7, S9, S10, S12, S13, and S15 switches used in leakage current testing as specified
in GB 9706.1-2020, which can be controlled via the device software. The instrument includes built-in test programs for leakage currents (including but not limited
to earth leakage current, contact current, patient leakage current, and patient auxiliary current) in accordance with GB 9706.1-2020, and allows selection of the
appropriate test program based on the type of applied part (Type B, BF, CF). Results for both normal condition and single fault condition can be obtained with one key.
4. Capable of testing with 110% voltage application, fully automatic testing, with storage capacity for data from at least 100 target devices.
5. Supplied with a complete set of test leads, alligator clips, and an enclosure probe.