Scope of Application
【Operating Instructions】
【Technical Parameters】
1. Challenge Pressure System
1.1 Three-channel pressure control system: independently controlled and independently detects
output pressure;1.2 Pressure Control Range: 10~400kPa;1.3 Pressure Display Accuracy: ±0.5% F.S.
(Full Scale);1.4 At least 3 units of 1-gallon pressure vessels, used for containing challenge bacterial
solution, wetting solution, and cleaning alcohol respectively;1.5 Pressure control mode: optional direct
pressurization or gradual pressurization; the amplitude and frequency of gradual pressurization are configurable;
2. Sample Challenge System
2.1 Sample Challenge Assemblies: At least 12 sets of sample challenge assemblies, compatible with clamping
samples of different sizes;2.2 Test Piping: Entire piping system can be autoclaved (high-temperature sterilizable);
2.3 At least 6 channels of adjustable all-metal flow control valves;
3. Test Environment
3.1 Biosafety Cabinet: Complies with the requirements of YY 0569-2011 "Class II Biosafety Cabinets";3.2 Average
Downward Air Velocity: 0.33±0.025m/s;3.3 Average Inlet Air Velocity: 0.53±0.025m/s;
【Configuration List】
1.Stainless steel pressure vessels (1-gallon): 3 units;
2.Aluminum busbar: 1 piece;
3.Test chambers: 12 units;
4.Air pipe: 3 meters;
5.Test chamber brackets: 2 pieces;
6.Air compressor: 1 unit;
7.Ultraviolet (UV) lamp tube: 1 piece;
8.Test chamber brackets: 2 pieces; (Note: Duplicate item retained as per original list; suggest verifying if it's a typo in the source text)
9.Biosafety cabinet: 1 unit.