Scope of Application
【Operating Instructions】
Applicable Scope:
This instrument is used to determine the ability of the covering material of diapers to resist the rewetting of previously infiltrated liquid back onto the skin.
Compliance Standards:
Complies with GB/T 24218.14 "Textiles - Test methods for nonwovens - Part 14: Determination of rewet of covering materials", ISO 9073-14 "Textiles - Test methods for nonwovens - Part 14: Determination of rewet of top sheets", WSP 80.10 - ERT 151.3(02), WSP 70.8 - ERT 154.0(02), and ISO reference 9073-14:2002.
Instrument Features:
Touch screen control with bilingual (Chinese and English) menus for convenient operation.
Built-in thermal printer for quick on-site printing of test results.
High-precision sensor with a test resolution of 0.001g.
Pneumatic components automatically lift the specimen when placing the simulated infant load, completing the test process automatically.
Controlled by a stepper motor and driver to ensure the speed complies with standard requirements.
The instrument features a built-in calibration function.
Technical Parameters:
Simulated Infant Load: Weight: 4000g ± 20g; Dimensions: 10 × 10 cm.
Test Range and Accuracy: 300g / 0.001g.
Test Speed: 5 cm / (5 ± 1) s.
Placement and Dwell Time: Adjustable from 1 to 999 seconds.
Tray Dimensions: 170 mm × 170 mm.
External Dimensions: 430 × 410 × 540 mm.
Weight: 51 kg.
Power Supply: AC 220V 50Hz 100W.