YS-227Y Medical Protective Clothing Synthetic Blood Penetration Tester (Single Station)
Search

YS-227Y Medical Protective Clothing Synthetic Blood Penetration Tester (Single Station)

Scope of Application

【Operating Instructions】

    Used for high-pressure penetration testing of test products by delivering high-pressure output.

**YS-227Y Medical Protective Clothing Synthetic Blood Penetration Tester (Single Station)**


**【Scope of Application】**  

Used for high-pressure penetration testing of test products by delivering high-pressure output.


**【Applicable Standards】**  

GB 4744-84  

ISO 811  


GB/T 19082-2009 Technical requirements for disposable medical protective clothing, etc.  


ASTM F1670 Standard test method for resistance of protective clothing materials to synthetic blood penetration  


ASTM F903 Standard test method for resistance of protective clothing materials to liquid penetration  


YY/T 0699-2008 Test method for resistance of protective clothing materials to penetration by pressurized liquids for liquid chemical protective equipment  


YY/T 0700-2008 Test method for resistance of protective clothing materials to penetration by blood and body fluids—Synthetic blood test method  


ASTM F1671 Standard test method for resistance of protective clothing materials to blood-borne pathogen penetration using Phi-X174 bacteriophage penetration as a test system  


ISO 16603:2004 Protective clothing against contact with blood and body fluids—Determination of the resistance of protective clothing materials to penetration by blood and body fluids—

Test method using synthetic blood  


YY/T 0689-2008 Test method for resistance of protective clothing materials to penetration by blood-borne pathogens—Phi-X174 bacteriophage test method  


ISO 16604:2004 Protective clothing against contact with blood and body fluids—Determination of resistance of protective clothing materials to penetration by blood-borne pathogens—

Test method using Phi-X174 bacteriophage  


**【Instrument Features】**  


1. Equipped with a silent air pump.  

2. Positive pressure sensor for real-time pressure monitoring.  

3. Precision pressure regulating valve with stepless adjustment, allowing setting of hold pressure, hold time, and other parameters.  

4. Adjustable spray time with automated operation for convenience and efficiency.  

5. Equipped with a mask clamping device for easy sampling and testing.


**【Technical Parameters】**  


1. Blood volume: 2 mL / 60 mL test  

2. Control mode: Manual / Push-button control  

3. Adjustable pressure up to 16 kPa / 2 psi for high-pressure output, accuracy 0.1 kPa  

4. Test pressure range: 0–16 kPa (2 psi)  

5. Micropressure gauge range: 0–0.1 kPa  

6. Pressure resolution: 0.1 kPa  

7. Time accuracy: 0.1 s  

8. Test medium: Synthetic blood (provided by customer)  

9. Control method: Manual / Push-button  

10. Blood volume: 2 mL / 60 mL


**【Test Procedure】**  


Secure the sample in the sample holder. At a distance of 30.5 cm from the sample center, spray 2 mL / 60 mL of synthetic blood (surface tension 0.042 ± 0.002 N/m) horizontally toward the target area of the sample through a needle with an inner diameter of 0.84 mm, at a pressure of 16 kPa / 2 psi. Remove the sample and visually inspect within 10 s / 60 s.