Scope of Application
【Operating Instructions】
【Scope of Application】Numerous instances indicate that bacteria can penetrate barrier materials via dry organic or inorganic particles. Examples include bacteria-carrying skin squames (flakes) penetrating surgical gowns or cleanroom suits, and bacteria-carrying particles penetrating packaging materials during storage.
The Dry Penetration Tester is used to determine the resistance of materials against bacterial penetration by dry particles within the size range of human skin squames.
【Compliance Standards】
ISO 22612: Clothing for protection against infectious agents — Test method for resistance to dry microbial penetration.
YY/T 0506.5-2009: Surgical drapes, gowns, and clean air suits for patients, medical staff, and instruments — Part 5: Test method for resistance to dry microbial penetration.
【Instrument Features】
Biological Cabinet: Equipped with a bio-cabinet that provides a sterile, dust-free purification work environment.
Lighting & Sterilization: Configured with internal lighting and ultraviolet (UV) sterilization devices.
Reliable Performance: Reliable product design, stable testing performance, and low noise (≤65dB).
User-Friendly Access: Features a front-mounted switch-type glass door for easy operation and observation by lab personnel.
Mobility: Detachable stand with 4 adjustable leveling feet; equipped with 4 auxiliary casters for easy movement.
Safety System: The biological cabinet includes a fan exhaust system; high-efficiency filtration devices at the inlet and outlet ensure operational safety.
Touch Interface: The tester adopts a full touch-screen interface for simple and convenient operation.
Easy Maintenance: The test container is made of all stainless steel, facilitating easy cleaning and maintenance.
【Technical Specifications】
1. Workbench
Vibration Form: Pneumatic ball vibrator
Vibration Frequency: 20,800 times/min (347Hz)
Vibration Force: 650N
Max Dust Concentration: 3500 particles/L (particle size ≥0.5μm)
Workbench Dimensions: 40cm × 40cm
Workbench Medium: Marble plate
Test Container: Stainless steel test container
Number of Stations: 6 stations
Filter Efficiency: Better than 99.99%
2. Biological Cabinet
Function: Provides air purification for a local dust-free and sterile work environment; controls the discharge of contaminated air from the work area through dedicated filtration channels.
Built-in Fluorescent Lamp: 18W × 3 units
UV Lamp: 20W × 1 unit
Cabinet Airflow Velocity: ≥0.3~0.5±0.025 m/s
Overall Dimensions: 1000mm (W) × 800mm (D) × 2040mm (H) (Upper and lower split type)
Weight: Approx. 190Kg
Air Source: ≥0.8Mpa
Power Supply: AC220V, 50Hz