Scope of Application
【Operating Instructions】
This instrument is used for testing breathing resistance. Its test method complies with the GB 2626-2019 standard. It features accurate and stable test data, as well as simple and convenient operation.
GB 2626-2019: Respiratory protection - Self-priming filtering particulate respirators (Clauses 6.5, 6.6).
GB/T 32610-2016: Technical specifications for daily protective masks.
GB 2890-2009: Respiratory protection - Self-priming filtering gas masks (Clause 6.9).
Resistance Measurement Range: -1000 ~ +1000 Pa.
Resistance Measurement Error: ±0.05%.
Test Flow Adjustment Range: 10 ~ 200 L/min.
Test Flow Stability Range: ±1 L/min.
Power Supply: AC 220 ± 10%, 50Hz.
Operating Conditions: Temperature 20℃ ± 2℃; Relative Humidity 65 ± 10%.
Durability: The instrument can operate continuously for 24 hours.
Dimensions: 530 × 400 × 800 mm.
Weight: 80 kg.