Scope of Application
【Operating Instructions】
Used for the detection of dead space in protective masks, specifically the volume fraction of CO2 in inhaled air.
GA 124: Clause "6.13.3 Determination of carbon dioxide content in inhaled gas" for Positive Pressure Firefighting Air Breathing Apparatus.
GB 2890: Clause "6.7 Mask dead space test" for Respiratory protection - Self-priming filter respirators against gases or vapours.
GB 19083: 5.5 Dead space for Medical protective masks.
GB 2626-2019: 6.9 Dead space for Respiratory protection - Self-priming filter respirators against particles.
GB 21976.7-2012: Part 7: Filtering self-rescue respirators for fire protection.
BS 149:2001 + A1:2009: 7.12 Carbon dioxide content of inhaled air.
Application Fields: PM2.5 masks, Medical protective mask product testing.
Operating Environment: CO2 concentration in ambient air ≤ 0.1%.
CO2 Gas Flowmeter: Range: 10-100 L/min; Resolution: 0.1 L/min; Accuracy: Class 0.5.
CO2 Gas Source: Volume fraction of CO2 gas is (5.0 ± 0.1)%. Resolution: 0.01%.
CO2 Gas Analyzer: Range: 0 ~ 12%; Repeatability Accuracy: Class I; Resolution: 0.1%.
Test Headforms: Available in three specifications: Large, Medium, and Small.
Handheld Anemometer: Range: 0-10 m/s; Accuracy: ±(0.1 m/s + 5% of reading) when measuring in the range of 0-2 m/s.
Fan Wind Speed: 0.5 m/s.
Artificial Lung: Breathing frequency adjustable range: (10-40) times/min; Tidal volume adjustable range: (0.5-3.0) L.
Power Supply: AC 220V, 50Hz.
Dimensions (L×W×H): 820 mm × 520 mm × 1380 mm.
Weight: Approx. 80 kg.