YS-610Y Medical Mask Flammability Tester
Search

YS-610Y Medical Mask Flammability Tester

Scope of Application

【Operating Instructions】

    The YS-610Y Flammability Performance Tester is mainly used to test the combustion performance of fabrics after contacting a flame at a specific linear velocity. It is a specialized testing instrument for fabric flammability performance.



YS-610Y Medical Mask Flammability Tester

Scope of Application

The YS-610Y Flammability Performance Tester is mainly used to test the combustion performance of fabrics after contacting a flame at a specific linear velocity. It is a specialized testing instrument for fabric flammability performance.

Compliance Standards

  • GB 19083-2010

  • YY 0469-2011

Instrument Features

  1. Fixture: Uses a metal human headform fixture, which fully simulates the actual usage state of a mask.

  2. Gas Connection: The burner is equipped with an interface for combustible gases such as propane (LPG).

  3. Adjustable Height: The burner height is adjustable.

  4. Automation: Automatic timed positioning of the burner.

  5. Data Recording: Both burning time and flame retardant time are automatically recorded and digitally displayed.

  6. Temperature Monitoring: Equipped with a flame temperature measuring probe and a flame temperature display.

Technical Specifications

  1. Headform Speed: The linear velocity of the headform nose movement is (60 ± 5) mm/s.

  2. Flame Height: Adjustable (Factory setting: 40 ± 4 mm).

  3. Probe Diameter: The diameter of the flame temperature measuring probe is 1.5 mm.

  4. Flame Temperature: At a height of (20 ± 2) mm from the top of the burner, the temperature is (800 ± 50) ℃.

  5. Burner Distance: The distance between the top of the burner and the lowest point is (20 ± 2) mm.

  6. Height Adjustment Range: Burner height adjustment range is 30 mm.

  7. Headform Specs: Metal headform wall thickness is 2~3 mm, resistant to flame and combustion.

  8. Timer: Afterflame time and afterglow time timer: 0 ~ 99.99 seconds; Accuracy: ± 0.01 seconds.

  9. Power Supply: AC 220V, 50Hz, 50W.

  10. Weight: Approx. 30 kg.



related product