YS-JF-3 Oxygen Index Tester
Search

YS-JF-3 Oxygen Index Tester

Scope of Application

【Operating Instructions】

    Suitable for determining the combustion performance of materials such as plastics, rubber, fibers, foam plastics, sheets, films, and textiles.

YS-JF-3 Oxygen Index Tester

【Scope of Application】
Suitable for determining the combustion performance of materials such as plastics, rubber, fibers, foam plastics, sheets, films, and textiles.

【Technical Specifications】

  1. Oxygen Sensor: Imported oxygen sensor with digital display; no manual calculation required.

    • Range: 0 — 100%

    • Accuracy: Higher precision and accuracy.

  2. Digital Resolution: ±0.1%

  3. Overall Measurement Accuracy: Class 0.4

  4. Flow Adjustment Range: 0 - 10 L/min (60 - 600 L/h)

  5. Response Time: < 5s

  6. Quartz Glass Cylinder: Inner Diameter ≥ 75mm, Height 480mm

  7. Gas Velocity in Combustion Cylinder: 40 ± 2 mm/s

  8. Flowmeter: 1 - 15 L/min (60 - 900 L/h) adjustable, Accuracy Class 2.5

  9. Test Environment:

    • Ambient Temperature: Room Temp ~ 40℃

    • Relative Humidity: ≤ 70%

  10. Input Pressure: 0.2 - 0.3 MPa (Note: Do not exceed this pressure).

  11. Working Pressure:

    • Nitrogen: 0.05 - 0.15 MPa

    • Oxygen: 0.05 - 0.15 MPa

    • Oxygen/Nitrogen Mixture Inlet: Includes pressure regulating valve, flow control valve, gas filter, and mixing chamber.

  12. Specimen Holders: Suitable for soft and rigid plastics, textiles, fire doors, etc.

  13. Ignition System: Propane (Butane) ignition system.

    • Flame Length: Adjustable freely from 5mm to 60mm.

  14. Gas Supply: Industrial Nitrogen and Oxygen, Purity > 99%.


related product
YS-JF-3 Oxygen Index TesterWhats the quotation? Message Us!